| |
BREG
Soil and Water Quality Lab
Quality Assurance and Quality Control Policy
1.0 INTRODUCTION TO LABORATORY
A. Mission Statement of Laboratory
The mission of the BREG Soil and Water Quality
Laboratory (BREG SWQL) is to provide accurate,
verifiable, affordable, and timely analytical
services. The laboratory is committed to quality
assurance, and it is our goal to provide high
quality project support and technical resources
to meet the needs of our department faculty as
well as other University of Delaware researchers.
The BREG Laboratory provides routine water and
soil analyses with special emphasis on nutrients
and water quality parameters.
B. Quality Assurance Policy of Laboratory
Dr.
Anastasia E. M. Chirnside will be responsible
for the hiring and training of staff with the
required qualifications and/or skills needed to
accomplish the mission of the laboratory.
The best available sample preparation, handling,
preservation and storage methods are used as recommended
by the appropriate authority. Procedures from
both Standard Methods for the Examination of Water
and Wastewater and the US EPA are used for water
preservation, storage, and analysis. AOAC (Association
of Official Analytical Chemists) recommendations
are used for agricultural and other types of materials.
It is recommended that the client discuss sampling
and preservation guidelines with the laboratory
personnel since these steps will affect the quality
of the data.
In addition to the above, Quality Assurance (QA)
also includes control of the following: calibration
and standardization, preventive and remedial maintenance,
proper instrument selection and use, quality laboratory
water, clean laboratory environment, replicate
analysis, spiking of samples, holding facilities
for samples, responsible evaluation of data, and
recording and maintaining a Quality Control (QC)
database. Examples of laboratory performance on
QC samples are available on request. Results of
QC samples are reported with the client's data
upon request.
C. Size of Laboratory
1. Dimensions and Layout
The BREG SWQL is housed in Worrilow Hall on the
University of Delaware campus. The primary laboratory
space is in Room 111 on the first floor of the
Worrilow Hall. Additional space is utilized in
Room 112 of Worrilow. A walk-in freezer and refrigerator
are located in room 122 of Worrilow Hall. Standard
refrigerators are located in both rooms 111 and
112. A chest freezer is located in Room 112. Plant
grinding equipment and soil grinding equipment
are available in the adjacent Soil Testing laboratory.
Sample driers and a large sample grinder are available
in the Agronomy Building located on the research
farm, which is adjacent to Worrilow Hall. Exhaust
hoods are available in both room 111 and 112 for
digestion and hazardous fume work.
2.0 LABORATORY ORGANIZATION AND RESPONSIBILITIES
A. Organizational Chart
The BREG SWQL is part of the Bioresources Engineering
Department in the College of Agriculture and Natural
Resources of the University of Delaware. The laboratory
is located on the Newark Campus in Worrilow Hall
at 531 S. College Ave., Newark, DE 19716-2140.
The laboratory provides analytical services primarily
to University researchers.
Department Head: Dr.
William F. Ritter
Laboratory Supervisor and Manager: Dr.
Anastasia E. M. Chirnside
B. Description of Lines of Communication
The manager of the BREG SWQL, Anastasia Chirnside,
reports to Bill Ritter, the Department Head of
the Bioresources Engineering Department. Dr. Chirnside
is responsible for accurate and timely analysis
of the materials submitted to the lab. This includes
all QA/QC procedures, equipment maintenance, hiring
and supervision of necessary personnel, updating
of equipment and procedures, organization of samples,
and scheduling of tests. The laboratory manager
is designated to be the QA\QC officer.
3.0 SAMPLE CUSTODY
A. Sample Receipt Policies
All samples are received in Room 111 of Worrilow
Hall on the Newark Campus of the University of
Delaware.
B. Sample Log-in
Once samples arrive at the laboratory, they are
recorded into the laboratory sample logbook (Excel
File). The logbook file contains the following
information: date received, storage location,
preservation required, number of samples in the
set, analyses required, and date completed. Each
project has a unique lab notebook. Each data set
name and lab numbers are recorded in designated
lab notebook. The lab number is used to identify
the sample in all analyses. After analysis is
complete, the samples are stored at 4º C
until the project investigators accept results.
Access to samples is limited to laboratory personnel.
For research samples, each sample set that is
logged into the lab is then assigned to a project
title and each sample is assigned a laboratory
number. Each project is defined as to the type
of analysis requested by the researcher.
C. Sample Storage and Preservation
If samples arrive in the laboratory unpreserved,
it is the responsibility of the section leader
to see that all preservation procedures necessary
to the individual analysis requested are performed
as soon as possible. Water samples are preserved
by refrigeration, freezing, or by the addition
of acid according to USEPA protocol. Soil and
plant samples are dried, ground and sieved in
accordance to NCR-13 and AOAC recommendations.
In instances where a soil sample is to be analyzed
for percent moisture or ammonium, the sample will
be refrigerated or frozen for storage.
D. Evidence Files
Each specific project is assigned a lab notebook
in which all log in sheets, notes, raw data, QA/QC
results, calculations used for data reduction,
and copies of reports produced by the laboratory
are recorded. Note books are kept in the lab for
at least 7 years.
E. Sample Disposal
Unless the client requests a return of their samples,
the laboratory is responsible for the disposal
of all samples. Samples are stored for at least
three months after the completion of analysis.
After this time, the samples will be subject to
disposal after approval by researcher.
If the samples are considered hazardous waste,
then disposal procedures follow the guidelines
set up by the University of Delaware Chemical
Waste Program, Department of Occupational Health
and Safety. Otherwise the samples are sewered
(water), composted (soil), or placed with other
refuse (plant and other).
If the client wishes to have either the samples
or the sample containers returned, those arrangements
must be made with the laboratory, preferably at
the time of the sample drop-off.
4.0 CALIBRATION PROCEDURES AND FREQUENCIES
A. Frequency of Calibration of All Instruments
The calibration of all instruments will be verified
at least once each day at the beginning of analysis
of unknown check samples. The calibration generally
includes at least one blank and several standards
bracketing the range of the samples. Some instruments,
like pH meters, do not require blanks, but use
several standards instead.
B. Labeling of Records of Calibration for Instruments
All records of the calibration of instruments
are kept in the laboratory analysis log book by
the section leader. Each record is labeled by
indicating date, analyses type, and analyst initials.
C. Reagents and Standards
1. Records of Receipt and Tracking
Records of dates ordered and dates received of
all commercially prepared check samples are recorded
in the QA/QC lab notebook. Additionally, if a
check sample needs to be diluted prior to use,
the preparation dates are kept on file by the
section leaders. Preparation of all lab standards
is recorded in the QA/QC notebook. Information
recorded includes: lot number of reagents, date
made, analyst performing task, and exact recipe
of how solutions were made.
2. Control Charts and Curves
Records of all raw data for control charts and
standard curves are recorded in the QA/QC notebook.
3. Disposal of Unused Standards
The disposal of unused standards follows the guidelines
established by the University of Minnesota Chemical
Waste Program, Department of Environmental Health
and Safety.
5.0 INTERNAL QUALITY CONTROL CHECKS
A. Internal Quality Control Checks:
In the laboratory, for each batch of 10 water
samples, a set of duplicate samples and a spiked
sample will be analyzed along with two standards;
one in the high concentration range and one in
the low concentration range. The standard and
spiked samples will be compared against previously
developed percent recovery data using the Shewhart
control chart. The duplicate analysis will be
compared to previous precision quality control
data using an R chart (EPA, 1979).
B. Performance and Systems Audits:
Quality control samples will be analyzed at the
beginning of the project and every 6 months to
evaluate measurement performance systems. Once
the project is initiated and on an annual basis;
all laboratory procedures will be checked and
evaluated. Checklists suggested in Procedures
for Evaluation of Environmental Monitoring Laboratories
will be used (EPA, 1978).
C. Preventive Maintenance:
All analytical balances are serviced on an annual
basis. In addition, the accuracy is evaluated
weekly by checking each balance with the appropriate
NIST weights.
D. Specific Routine Procedures Used to Access
Data Precision, Accuracy, Representative- ness
and Completeness.
Data accuracy will be checked for each set of
samples by comparing the spiked and standard recoveries
against a Shewhart control chart. Precision will
be checked on duplicate samples for each set of
analysis using an R chart. The equations and procedures
for checking precision and accuracy are presented
in the EPA Manual on Handbook for Analytical Quality
Control in Water and Wastewater Laboratories (EPA,
1979). Representativeness of the analysis will
be checked by comparing individual data sets with
mean concentrations and standard deviations established
from previous analysis. Completeness of the data
set will be checked by determining what percent
of data is rejected from data sets based on all
data collected.
E. Laboratory Water Purity
The water used for all analytical purposes in
the laboratory is triple deionized to a minimum
resistivity of 17.5 megohm. Worrilow Hall, where
the laboratory is located, has a building reverse
osmosis system t produces 2 megohm DI water. This
is piped to the various laboratory rooms for general
use. In the BREG SWQL, the water is piped to a
Barnstead E-Pure system which uses macroreticular
resins as an initial step to remove colloids,
activated carbon as a second step to remove organics
and chlorine, two ultrapure mixed bed cartridges
to remove all ionic contaminants, and finally
a 0.2 micron cross flow filter in a remote dispenser.
This process produces water which is continually
digitally monitored to be at least 17.5 megohms,
and meets the requirements of Type I water as
determined by the ASTM (American Society for Testing
and Materials).
F. Reagent Storage and Purity
Chemicals and reagents used in the laboratory
are Certified ACS grade as required in the SOPs.
For trace metal digestions and ICP-AES analysis,
the acids are TraceMetal grade which are manufactured
to achieve low metal contamination measurable
in the mg/L to u/L range.
All prepared reagents used in the laboratory are
stored according to the specifications prescribed
in the SOP, i.e. brown bottles, glass only, under
refrigeration etc. Only the amount of reagent
necessary for each day's run is removed from the
stock bottle. This helps eliminate the possibility
of contamination of the stock reagent and also
allows for continuous refrigeration of those reagents
for which that is a concern. The purity of the
reagents is ensured by the laboratory's procedures
for glassware cleaning, water purity, and technical
procedures in the preparation. All these are designed
to protect against contamination. The purity is
protected by the policy of only pouring out enough
reagent for each day's run, never withdrawing
reagent from the original container with any type
of apparatus, never pouring excess reagent back
into the container, and following as closely as
possible all conditions for storage.
G. Bottle Cleaning
All glass and plastic ware are to be washed with
nonphosphate detergents and thoroughly rinsed
(see glassware cleaning appendix). Some tests
call for an appropriate acid rinse as outlined
in the SOP for that test. For most procedures,
it is recommended to use newly purchased containers.
This eliminates any concern of contamination from
a previous sample. All glassware is then rinsed
with chromerge (acid rinse). The plastic ware
and plastic bottles are treated by a double acid
rinse of 5% Nitric acid followed by 5% Hydrochloric
acid. After the acid rinse step, they are thoroughly
rinsed with deionized water (at least three final
rinses with 18 megohm deionized water), and air
dried upside down.
6.0 DATA REDUCTION, VALIDATION, AND REPORTING
A. Procedures of Rerunning Data
Samples are rerun when they are associated with
a QC check sample that does not fall within the
95% confidence interval of the expected value.
All of the samples following the last valid QC
check are rerun. If the QC check sample is still
outside the 95% confidence interval or if the
associated check sample does not fall within 3
standard deviations of the historical or established
mean, the procedure is considered out of control.
These events initiate trouble shooting and corrective
action procedures. The QC check sample is rerun
and validated before reruns of the unknowns resume.
B. Procedures for Flagging Data
Data is flagged when there is a deviation from
the established SOP or QA/QC criteria. This action
is communicated to the lab manager by the analyst
performing the analysis. The deviation that caused
the flagging and any resultant corrective actions
are discussed with the laboratory manager. Samples
are generally rerun following the corrective action,
but in cases where no corrective action can resolve
the problem (i.e. a holding time was missed, the
sample is depleted and cannot be rerun, or the
standard addition procedure was used for calibration
etc.), then the data on the final customer report
is flagged and an explanation is given on the
report which notifies the customer of the deviation.
C. Use of Performance Evaluation Standards
Performance evaluation standards are used to evaluate
the method, the SOP, and the performance of the
analyst. If the performance evaluation of the
laboratory is not acceptable to the reviewing
agency, then trouble shooting and corrective action
is initiated. The validation resulting from the
successful analysis of performance evaluation
standards lends confidence to the methods, procedures,
and analysis of the laboratory. The performance
evaluation groups in which this laboratory participates
are listed in part J below.
D. Practical Quantitative Limits
Practical method detection limits (see section
9.0 part E) are established for each analysis
in the laboratory. Result values that fall below
the MDL will be reported as less than the MDL
value.
7.0 ROUTINE PROCEDURES TO ACCESS DATA QUALITY
AND DETERMINE REPORTING LIMITS
A. Precision
Precision is the agreement among a set of replicate
measurements without the assumption of knowledge
of their true values. There are two primary means
of evaluating precision in this laboratory. The
best mechanism to evaluate precision is the examination
of relative percent difference of duplicate samples
in the analytical run. This is expressed in the
formula:
RPD = 100[(X1-X2)/{(X1+X2)/2}]
Where RPD = Relative Percent Difference
X1 = First observation of unknown X
X2 = Second observation of unknown X
Sample unknowns are duplicated at the rate of
one per every 10-20 unknowns, depending on the
SOP. Relative percent differences of 10% are expected
at levels of ten times the MDL and above. When
unacceptable RPD values are encountered, the associated
data (a batch of 10 to 20 samples) is rerun after
a QA review. As the analyte level approaches the
MDL, 10% RPD is too strict and higher RPDs are
acceptable. At these low levels, the RPD is evaluated
with reference to the PQL and other QC data in
the run.
Another mechanism to evaluate precision involves
a comparison of a check sample run daily with
each batch of samples. If the check sample is
run several times during the analytical run, then
an estimate of replicability of the run can be
obtained. The standard deviation of these results
is an estimate of daily precision. The repeatability
of the SOP over time can be evaluated by the comparison
of the results of this check sample on a day to
day basis. The pooled standard deviation of the
check sample over many days and analyses gives
an evaluation of the precision of the method over
time.
B. Accuracy
Accuracy is the measure of bias of an analytical
procedure which reflects the closeness of a measured
value to a true value. In this laboratory, accuracy
is daily measured on water, soil, compost and
plant samples for those tests where certified
values are available. Standard reference materials
(SRM) of soil, sediments, and plants containing
certified levels of analytes are purchased from
US NIST for use as control checks on accuracy.
Certified water control samples are prepared by
dilution of water standards obtained from SPEX
Industries, Inc., Edison, NJ and Perkin Elmer
Corp., Norwalk CT.
For water analysis, at least one check control
sample containing the appropriate certified analytes
is run in conjunction of each batch sample unknowns.
For other analyses, a certified or in-house control
check sample is included in each run. For large
sample batches, one check sample is run for each
group of 20 unknowns. A batch of samples is rerun
when the observed value for the associated check
falls outside of the expected two standard deviation
confidence interval. The analysis is considered
out of control if the observed check values are
not within three standard deviations of the known
mean. In this case trouble shooting and corrective
action is initiated.
In addition, no more than seven consecutive control
check values should fall on one side of the historical
mean, even if they are within the acceptable confidence
limits. Such an event would be evidence that there
is drift or bias in the procedure, and initiates
a QC evaluation.
C. Representativeness
Representativeness is the assurance that the sample
or subsample used in the laboratory is indeed
representative of the field entity that is being
measured. There are two major stages involved
in representativeness. One is the collection of
the sample from the entity being measured, and
the other is sample preparation and homogenization
before the sample is subsampled by the laboratory.
Sample collection is not a service offered by
the laboratory except in very special circumstances.
However, to help ensure that the sample is collected
in a manner that insures it is representative
of the field entity, the laboratory will offer
to a client upon request any information we might
have on sampling techniques and preservation.
The conditions applied to the actual sampling
procedure, the decision as to how to preserve
the samples, the interim storage, and the means
of delivering the samples to the laboratory are
the responsibility of the client.
Sample preparation and subsampling services are
provided by the laboratory. Each sample submitted
must be prepared and homogenized to ensure that
any subsequent subsample taken by the laboratory
is representative of the sample originally submitted
by the client. This can include tasks such as
drying, grinding, sieving and mixing of the sample,
depending on its composition. Plant samples are
ground to pass a 20 mesh sieve and soil samples
are crushed and sieved to pass a 10 mesh sieve.
All samples are shaken or stirred before subsampling.
For water samples with significant sediment, the
client is consulted for instruction. Water samples
are not filtered unless specified by the client.
Water samples are thoroughly mixed before a subsample
is withdrawn.
D. Completeness
It is the responsibility of each laboratory manager
to review the analytical report for each job to
insure that (1) all the samples required for quality
assurance and quality control have been processed
and (2) that the analytical report to be retained
on file contains a complete record for each analysis
and the associated QC samples and (3) that all
procedures specified by established QA/QC protocols
have been implemented.
E. Reporting Limits
1. IDLs
The instrument detection limit is the smallest
signal above the background noise that the instrument
can reliably detect. For most methods this would
be a concentration of analyte that produces a
signal greater than five times the signal/noise
ratio for that instrument.
2. MDLs
The method detection limit is the analyte concentration
derived from the method that yields a signal which
is large enough to be considered significantly
different from the blank with a statistical 99%
probability. The method detection limit is determined
by analyzing reagent water fortified at a concentration
considered to be two to three times the estimated
detection limit. At least seven replicates of
this fortified blank are analyzed by the same
procedure followed in the determination of unknown
samples. The MDL is then calculated using the
equation MSDL = (t) x (S), where t = 3.14 (for
seven replicates) and S = the standard deviation
of the replicate analysis. Upon request of the
client, this value can be used as a reporting
limit.
3. MDLs
The method detection limit is the lowest value
that can be arrived at reliably during normal
routine laboratory analysis. For most analyses,
this is specified as three times the MSDL. This
value is routinely used as the reporting limit
for unknowns reported to the client.
8.0 CORRECTIVE ACTION
When quality control observations fall outside
established acceptance criteria in terms of precision
and accuracy, and continue to fail with reanalysis
within SOP, the procedures are reviewed by the
Laboratory Manager. More specifically, the following
actions might be taken: (1) standards and samples
are rerun to check if the instrument is running
properly and operating conditions are stable,
(2) new standards are prepared from a second stock
solution and run to check the original standards,
(3) the equipment is recalibrated and (4) new
reagents are prepared. The exact corrective action
conducted by the laboratory section leader will
vary depending upon their observation and experience,
e.g. whether the issue is one of precision or
accuracy. The action taken is recorded in the
log book accompanying each instrument.
9.0 QUALITY ASSURANCE REPORTS TO MANAGEMENT
Detailed quality assurance reports are available
upon request. The report could include factors
such as (1) changes or modifications to the QA/QC
Plan, (2) changes to the SOP in methods performed
for the client, (3) significant QA/QC problems
and the recommended solutions, (4) corrective
actions taken and the results, (5) limits that
shall be imposed on the data, (6) holding times
that have been missed, (7) recent management or
personnel changes that may have affected the work,
and (8) other issues that may have affected the
analysis. Clients in ongoing projects are notified
if major changes, such as new methods, different
instrumentation, personnel changes etc. take place.
e |
|
|